logo

산업 가스 제품군에 대한 GnL 솔루션 정보를 목록화 하였으며, 해당 솔루션에 대한 정보를 빠르고 쉽게 이용 하실 수 있습니다.

MTO 1000 H2O

2024-06-25
MTO-1000-H2O.pdf [585.83 KB]



MTO 1000 H2O

Trace Moisture Analyzer for Inert, Passive, Toxic and Corrosive Gases
--------------------------------------------------------------------------------

[View Small] [View Medium] The MTO-1000 H2O moisture analyzer covers a wide dynamic range, from parts per trillion (PPT) to parts per million (PPM), with unmatched accuracy, reliability, speed of response and ease of operation.

The MTO-1000 H2O analyzer features:

Tiger Optics’ patented Cavity Ring-Down Spectroscopy-based methane sensor
Optional 4-inlet sample manifold and internal valve actuation control to switch between 4 sample lines
Capability to measure moisture in most inert, corrosive and toxic gases with just one device
Totally automated maintenance. Users enjoy freedom from typical analyzer requirements such as periodic sensor maintenance, span calibrations, reliance on permeation devices, purifier replacement and use of vacuum pumps Mass flow controller to control flow rate of sample gas

MTO 1000 H2O -- Specifications

Gas Sample Conditions
Sample Inlet Pressure: 10 - 125 psi (1.7 - 9.6 bar)
Flow Rate: Up to 2 slm per inlet
Sample Gases: MTO-1000 Inert and passive gases
MTO-1001 Inert, toxic, corrosive, and passive gases
Sample Line Temperature: Up to 60°C

Gas Flow System
Materials of Construction: 316L stainless steel (optional Hastelloy?)
Wetted Components: 10 Ra surface finish
Gas Connections: Sample inlets & outlets - 1/4" VCR™
Automatic Sampling System: MTO-1000 One input
MTO-1001 Four inputs standard
Leak Tested to: <2 x 10-8 mbar • l /sec
Pneumatic Pressure (MTO-1001): 70 - 100 psig, clean dry air
1/8" Press-to-lock

Electrical
Alarm Indicators: User-programmable alarm set points
Power Requirements: 90-240 VAC 50/60 Hz
Power Consumption: 200 Watts max.
Output Signals: Recorder Isolated 4-20mA or 0-20 mA output per sample inlet
Alarm Form-C relay per sample inlet
Form-C relay for system fault
(Laser status, flow status, ring-down status)
User Interface: 10.4" color VGA display with touch screen, PS2 Mouse and Keyboard connection, 10BaseT Ethernet, USB, RS-232

Dimensions
Mounting: 14" x 19" x 29" (356mm x 483mm x 737mm)
Weight: 100 pounds (45kg)

Performance
Lowest Detection Limit: 200 ppt
Sensitivity 100 ppt
Accuracy (greater of) 4% of reading or +/- 0.1 ppb
Speed of Response (typical): Response to 50 ppb intrusion - less than three minutes
Operating Range: 0 - 5000 ppb. Customized higher ranges are available.
Environmental Conditions: 10° - 40°C
Storage Temperature: -10° - 50°C

Technology
Approvals: CE: LVD & EMC approved
Method: Cavity Ring-Down Spectroscopy
Patents: U.S. Patent #5,528,040
Other Patents Pending